FDA Flags Overheating Risk in Power Cords of Tablo Dialysis Machines

FDA Flags Overheating Risk in Power Cords of Tablo Dialysis Machines

The FDA has issued an Early Alert after Outset Medical notified customers that overheating can occur at the connection between the power cord and power inlet on its Tablo Console and TabloCart with Prefiltration Drawer. The company sent a letter on November 10 recommending updated use instructions rather than a device replacement, and no serious injuries or deaths have been reported as of that date.

What Triggers the Overheating

According to Outset Medical, the problem arises when the power cord connector is not held firmly in the power inlet. This loose connection allows the connector to shift, which increases electrical resistance and generates heat at the junction. The issue has been identified in both versions of the cord: the original clear-ended cord (PN-0005746) and the current gray-ended cord (PN-0008460), meaning the design change between versions did not eliminate the risk.

Warning Signs Facilities Should Watch For

The FDA and Outset Medical outlined specific indicators that a power cord connection may be overheating, including a burning or hot-plastic smell, visible smoke, a warm or hot cord or connector, melting or discoloration at the connector or inlet, sparks, flames, or unusual electrical sounds such as popping or sizzling. Additional signs include a loose-fitting connector, cracking or fraying along the cord, a tripped circuit breaker, sudden loss of power to the console, or a power-related alarm. The agency notes that a person touching an overheated connector could suffer a first- or second-degree burn, and that lost power could delay, interrupt or cause a missed dialysis treatment.

Recommended Precautions for Users

Outset Medical is not asking facilities to stop using the Tablo systems. Instead, the company is advising continued use under stricter precautions:

  • Inspect the power cord and connection before every use for warmth, looseness, cracking, fraying, melting or discoloration.
  • Ensure the cord is fully and firmly inserted into both the console (and TabloCart, if used) and the wall outlet.
  • Avoid pulling, bending or straining the cord, and keep it away from walkways where it could be caught.
  • Use only the power cord supplied by Outset Medical - no extension cords or power strips.
  • If overheating signs appear, stop use, follow facility shutdown procedures, unplug the unit once safe, and contact Outset Technical Support before returning it to service.

Facilities are also asked to share the notice with all staff who operate or maintain the devices, forward it to any organization that received a transferred unit, and inform home dialysis patients if applicable.

Regulatory Status and Reporting

The Tablo Hemodialysis System is cleared for use in patients with acute or chronic renal failure, with or without ultrafiltration, in both acute and chronic care settings. The FDA's Center for Devices and Radiological Health (CDRH) describes this as a potentially high-risk device issue and says it will update its public alert page as new information emerges. Health care professionals and consumers can report adverse events or quality problems through the FDA's MedWatch program, and U.S. customers with questions may contact Outset Medical directly.